The Taigan SL gas sampling system is an automatic, digital, one - or two-channel continuous device designed to measure the dew point temperature and the volume fraction of oxygen in nitrogen, hydrogen, inert and other non-aggressive gases. Provides visualization and registration of measurement results, as well as the output of relevant data to external monitoring and monitoring devices.

  • Characteristics
  • Applications
  • Download

Depending on the tasks being implemented, the device can be presented in the following modifications:

"Taigan 01 SL" - measurement of the dew point temperature;
"Taigan 02 SL" - measurement of oxygen content;
"Taigan 03 SL" - measurement of dew point temperature and oxygen content.

Technical specifications:
Number of analytical channels in the system, pcs. 1 or 2;
Maximum allowable gas pressure:

at the sampling point, MPa . . . . . . . . . . . . . . . . . . . . . ≤ 1;
at the point of sample submission for analysis, MPa * . . . . . . . . . . 0,06;
Measured range of gas flow, l / min . . . . . . . . . . 0,2 ~ 10;
Adjustable gas flow range, l / min . . . . . . . 0 ~ 10,5;
Measurement range of oxygen impurity, ppm **. . . 0 ~ 10 000;
Measuring range of dew point temperature, °C... -80 ~ -20;
Operating conditions of the system application:

temperature, °C . . . . . . . . . . . . . . . . . . . . . . . . . . 5,0 ~ 35;
relative humidity, % . . . . . . . . . . . . . . . . . . to 80;
atmospheric pressure, kPa . . . . . . . . . . . . . . . 84 ~ 106,7.

* For Taigan 01 SL, the maximum allowable gas pressure at the sample delivery point for analysis is 0.9 MPa.
** It is possible to measure the oxygen content in eight percent ranges (1, 2, 5, 10, 21, 25, 50 and 100 %). All ranges are user-selectable.

The device is available as part of a desktop or mounted unit.
The measurement of impurity concentrations in the analyzed gas stream is carried out by the forced supply of gas from the controlled zone to the sampling system.

As a rule, the system uses a direct sampling method. In this case, the gas sample is removed from the flow and directly sent through the sampling line to the input of the analytical units of the system.

However, if the gas flow parameters correspond to the technical characteristics of the system, then it can be installed sequentially, directly into the main line of this gas flow.

Purpose:

The gas sampling system "Taigan SL" is an automatic, digital, one - or two-channel continuous device designed to measure the dew point temperature and the volume fraction of oxygen in nitrogen, hydrogen, inert and other non-aggressive gases. Provides visualization and registration of measurement results, as well as the output of relevant data to external monitoring and monitoring devices.